InstructionsandOperating ManualX96SWEIGH SCALE60%Ronan EngineeringMain MenuVariablesStatus DisplayConfigurationDigital OutputsDigital InputsAuto CalCa
6PasswordNotice:To access the Programming Menu, the Password is 101010.Step 1: Power Up – You should now be on the Status Screen.Step 2: Press F3 to g
7Menus/OperationMenu TreesThe Ronan X96S Weigh scale uses a tree structured menu system.Ronan X96S – Weigh ScaleVariables VariablesStatus DisplayConfi
8ConfigurationOperation FilteringOperationRate Filtering Dyn Track [enable]Weight Detector Fault Sigma [number]Speed Linearization Fast TC [seconds]He
9Continued fromPrevious PageConfigurationOperation Alarms Alarm [number]RateWeight Alarm 1 Source [variable]Speed Alarm 2Alarm Type [none,low,high,ran
10Digital OutputsOutput OutputSelect SourcesPolarity Relay 1Relay 2Relay 3Relay 4TTL 1TTL 2TTL 3TTL 4Select SourcesAlarm 1 [yes/no]Alarm 2 [yes/no]Ala
11CalibrationState [state]Ref Constants Ref Constants Low ReferenceCalibrateCal Speed Ref Mode [mode] ReferenceRef Date [date] Ref Time [time] Ref Wei
12Root MenuThe root menu is titled “Ronan X96S – Weight”. It contains the following items:ITEM FUNCTIONVariablesSelecting this choice takes the user t
13Displays MenuThe menu titled “Displays” contains the following item:ITEM FUNCTIONStatus DisplaySelecting this choice takes the user to the Status Di
14Configuration MenuThe Variables menu is used to access area configuration menus. It contains the following items:ITEM FUNCTIONOperationSelecting thi
15Filtering MenuThe Filtering menu is used to configure the parameters associated with the weight measurement filter. It contains thefollowing items:I
iTable of ContentsOVERVIEW...1Ad
16Linearization MenuThe X96S is capable of performing a multi-point linearization of the weight data when required by an application. Thelinearizatio
17Rate MenuThe Rate Config menu is used to configure the parameters associated with the Rate measurement. It contains the followingitems:ITEM FUNCTION
18Weight MenuThe Weight Config menu is used to configure the parameters associated with the Weight measurement. It contains thefollowing items:ITEM FU
19Head Temp Config MenuThe Head Temp Config menu is used to configure the parameters associated with the detector electronics temperaturemeasurement.
20Remote Totalizer MenuThe Remote Totalizer menu is used to configure the parameters associated with the Remote Totalizer. It contains thefollowing it
21AlarmsThe Alarms menu is used to configure the parameters associated with the analog alarms.ITEM FUNCTIONSourceShow, and allows the user to set the
22Hardware MenuThe Hardware menu is used to define the type of hardware used to provide measurements and radiation. It contains thefollowing items:ITE
23Source Type MenuThe Source Type menu is used to define the type of radiation source used. It contains the following items:ITEM FUNCTIONSource TypeSh
24Analog Out Cnfg MenuThe Analog Out Cnfg menu is used to set where the source of power is internal or external. It contains the following items:ITEM
25Digital Outputs MenuThis menu is used to view and configure the digital outputs. It contains the following items:ITEM FUNCTIONOutputShows and allow
iiSpeed Hardware Menu...
26Digital Inputs MenuThis menu is used to view and configure the digital inputs. It contains the following item:ITEM FUNCTIONInput 1Selecting this it
27Calibration MenuThis menu is used to view and control the calibration of the X96S Weigh Scale. It contains the following items:ITEM FUNCTIONStateSh
28Calibrate MenuThis menu is used to access the various weight calibration procedures. It contains the following items:ITEM FUNCTIONStateShows the st
29ITEM FUNCTIONLow Speed CalThis item invokes a method that performs the low speed calibration procedureLow SpeedShows and allows the user to set the
30Aux Loop Cfg MenuThis menu is used to access the secondary 4-20ma loop calibration procedures. It contains the following items:ITEM FUNCTIONSV isSh
31X96S Local DisplayThe X96S Local Display consists of a 16 line by 21 characterdisplay and a 10 key keypad. The top line of the display isreserved fo
32Editing ValuesThe editing of different types of values is designed around the use of the four direction keys and up to 4 function keys. The leftand
33SourceInspection andInstallationSafety PrecautionsPlease see the Basic Radiation Safety Manual which isshipped with the source and source holder. I
34Mechanical MountingReview the Configuration Drawing which isincluded in the Drawing Chapter of this manual.Please reference the dimensional drawings
35Electrical Installation of Interconnect WiringDO NOT APPLY POWER until wiring is carefullychecked.Wire the equipment according to the detailedinterc
iiiDETECTOR ...49Scintillator Detec
36Microprocessor VerificationRotate latch clockwise to open the enclosure door.Next remove the computer front cover by sliding theblack tabs down. Che
37Power-upBefore applying power, ensure all boards are fullyseated in frame’s slots. Close front door of the X96Sand secure the door...When power is
38PasswordNotice:To access the Programming Menu, the Password is 101010.Step 1: Power Up – You should now be on the Status Screen.Step 2: Press F3 to
39CalibrationInitial calibration consists of two parts, the belt speed calibration and the mass density calibration.The belt speed calibration correla
40High Speed CalibrationPROCEDURE(For weigh scales using tachometer or encoder for speed). If speed input is a signal from a line downcontact, see Spe
41Speed Calibration with Line Down ContactPROCEDURE(For weigh scales using line down contact for speed signal)(If the weigh scale uses encoder or tach
42Mass Density Low/Reference and High/CalibrationThe mass density calibration correlates the detector counts to the mass upon belt. This can be achiev
43High CalibratePROCEDURE (Option 1)Passing or collecting a known amount of material through the scale and manually entering the L.F.value(Loading Fac
44High Calibrate (Continued)PROCEDURE (Option 1 Continued)Calculate the new L.F. (Loading Factor) and input the value into the X96SUsing the Ronan loc
45High Calibrate (Continued)PROCEDURE (Option 2)Placing a static load on the belt that you must weigh):Before proceeding, ensure that the belt is load
1OverviewThe X96S is a family of measurement products that is intended to replace the current X96N and X99 product families. Theseproducts:• use nucl
46High Calibrate (Continued)PROCEDURE (Option 3)An alternative to calibrating with process material is to use an absorber plate. The procedure is muc
47For future reference, document these items:(a) Environmental/process conditions (densities, temperatures, etc.) that influence the reference/cali
48ConfigurationRonan ships the Weight Scale System with factory-default software settings. Those settings are responsible for the informationthat ini
49DetectorScintillator DetectorDescription The Ronan scintillation detector consists of three main components: Theplastic scintillation crystal, the p
50Detector Service The critical components of the electronic circuit and thePMT/Crystal Assembly are aligned before leaving thefactory. If any compon
51ION ChamberDetector/Amplifier Ronan’s ion chamber detector is filled with an inertAssembly high-pressure gas. It uses low-voltage (-15VDC) biasand
52Servicing the Detector The ion-chamber detector contains pressurized inert gas. The ionchamber itself is not serviceable and must be returned to th
53Detector Removal/Replacement\ELONGATED DETECTORS NOTES:To avoid damage in shipment or installation, theelongated detectors are packaged separate fro
54Removing the DetectorAmplifier Circuit Board(CBAY-6102)Follow this procedure to remove the electrometer amplifier circuitboard:1. Remove the amplif
55ElectronicsX96-2001PLX96-2001PL is a CPU Module that comes without firmware.X96-2002PLX96-2002PL is the local Graphics L.C.D. Module. This optional
2Basic ConceptsCommunicationsThe Ronan X96S Weigh Scale provides both 4-20 mA current loop and HART communications.4-20 MAFor many years, the field co
56X96-2008PLX96-2008PL is the Digital Input/Output Module. A total of 16 bits of digital I/O and wetting/encoder power is provided by themodule.8 isol
57OptionsX96S Mechanical Chassis Part NumbersPART NUMBER DESCRIPTIONX96S-N4-1 X96S NEMA 4 Enclosure, 6 Position, with Front Panel L.C.D. & Motherb
58SPECIFICATIONSMODEL X96SProcess Computer: Microprocessor-based unit wit a liquid crystal display, push-button interface,HART® Communications, proces
59RegulationsRegulations will be supplied with Radiation Safety Manual.
60Drawings5.875[149.21mm]11.500[292.11mm]2.000[50.80mm]7.500[190.50mm]11.500[292.10mm]9.500[241.30mm]0.312[07.92mm]X96-N4-1Outline AssemblyDrawing
61AB8.000[203.20mm]10.340[262.64mm]X96S-SMOutline/AssemblyDrawing
62Mounting Plate
63X96STerminal Arrangements
64RONAN ENGINEERING COMPANY21200 Oxnard StreetWoodland Hills, California 91367 U.S.A.(800) 327-6626 • FAX (818) 992-6435E-Mail: [email protected] Sit
3Device Variable ValueRate RateWeight WeightSpeed SpeedHead Temp Head TemperatureConfiguration VariablesThe Ronan X96S Weigh scale has many configurat
4TheoryTheory of Radiation GagingRadiation gages operate on the principle of radiation absorption and transmission.A beam of gamma radiation is direct
5Principles of OperationThe detector's raw output signal is processed through several stages of software in the X96S.Some of the more significant